Related Experiment Video
Updated: Apr 30, 2026

Elemental-sensitive Detection of the Chemistry in Batteries through Soft X-ray Absorption Spectroscopy and Resonant Inelastic X-ray Scattering
Published on: April 17, 2018
An ultra-high vacuum electrochemical flow cell for in situ/operando soft X-ray spectroscopy study
Debajeet K Bora1, Per-Anders Glans1, John Pepper1
1Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, California 94720, USA.
Abstract:
An in situ flow electrochemical cell has been designed and fabricated to allow better seal under UHV chamber thus to achieve a good signal to noise ratio in fluorescence yield detection of X-ray absorption spectra for spectroelectrochemical study. The cell also stabilizes the thin silicon nitride membrane window in an effective manner so that the liquid cell remains intact during X-ray absorption experiments. With the improved design of the liquid cell, electrochemical experiments such as cyclic voltammetry have been performed for 10 cycles with a good stability of sample window. Also an operando electrochemical experiment during photoelectrochemistry has been performed on n-type hematite electrode deposited on silicon nitride window. The experiment allows us to observe the formation of two extra electronic transitions before pre edge of O K-edge spectra.