Double threshold ultrasonic distance measurement technique and its application.
Weihua Li1, Qiang Chen1, Jiangtao Wu1
1Key Laboratory of Thermo-Fluid Science and Engineering, Ministry of Education, Xi'an Jiaotong University, Xi'an, Shaanxi 710049, China.
The Review of Scientific Instruments
|May 3, 2014
Summary
A novel double threshold method accurately measures ultrasonic time-of-flight (TOF), suppressing noise for precise distance and gas flow measurements. This low-cost, real-time system offers high accuracy with minimal deviation.
Area of Science:
- Instrumentation and Measurement
- Acoustics and Ultrasonics
- Signal Processing
Background:
- Accurate time-of-flight (TOF) measurement is crucial for ultrasonic applications.
- Existing TOF techniques face challenges with signal noise and real-time processing.
- Developing cost-effective and precise ultrasonic measurement systems is an ongoing need.
Purpose of the Study:
- To adapt and validate a double threshold method for accurate ultrasonic time-of-flight (TOF) measurement.
- To develop a robust system for ultrasonic distance and gas flow measurements.
- To demonstrate the method's effectiveness in noise suppression and temperature/pressure compensation.
Main Methods:
- Implementation of a double threshold method using hardware circuits and the TDC-GP21 timing chip.
- Development of high-precision temperature (10 mK) and pressure (0.01% accuracy) measurement modules.
- Experimental validation of the system for both ultrasonic distance and gas flow measurements.
Main Results:
- Achieved a time resolution of approximately 22 ps with effective noise suppression.
- Demonstrated low cost and ease of implementation for the double threshold method.
- Distance measurement yielded maximum absolute deviation of 0.69 mm (0.28% relative error) from 100-600 mm.
- Gas flow measurement showed maximum absolute deviation of 1.16% and 0.65% repeatability for 50-700 m³/h flow.
Conclusions:
- The double threshold method is a viable, low-cost solution for accurate ultrasonic TOF measurement.
- The developed system reliably measures distance and gas flow with high precision.
- The system's hardware requires no adjustment for use as a two-path ultrasonic gas flowmeter.
Related Concept Videos
Electronic Distance Measuring Instruments
771
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over...
771
Design Example: Measuring Distance Between Two Points with Obstructions
581
When measuring distances in areas with physical obstructions, such as a lake in a field, surveyors must employ techniques to calculate accurate lengths without direct line measurements. One effective method is the offset technique, which allows for precise distance estimation over inaccessible stretches.In this scenario, a surveyor must measure a side of an area that crosses a lake. Since the measuring tape cannot span the lake, the surveyor begins by establishing a baseline that aligns with...
581
Ultrasonography
6.5K
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
During an ultrasonography procedure, a handheld device called...
6.5K
Echo
1.2K
The human ear cannot distinguish between two sources of sound if they happen to reach within a specific time interval, typically 0.1 seconds apart. More than this, and they are perceived as separate sources.
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
1.2K