[Evaluation of transit in vivo dosimetry using portal imaging and comparison with measurements using diodes]
P Royer1, V Marchesi2, V Rousseau3
1Département de radiothérapie, institut de cancérologie de Lorraine Alexis-Vautrin, 6, avenue de Bourgogne, 54500 Nancy, France; UMR 7039, centre de recherche en automatique de Nancy (Cran), BP 70239, 54506 Vandœuvre-lès-Nancy cedex, France; UMR 7039, université de Lorraine, BP 70239, 54506 Vandœuvre-lès-Nancy cedex, France; UMR 7039, CNRS, BP 70239, 54506 Vandœuvre-lès-Nancy cedex, France.
Purpose:
In vivo dosimetry transit using portal imaging is a promising approach for quality assurance in radiotherapy. A comparative evaluation was conducted between a commercial solution, EPIgray(®) and an in vivo dosimetry control reference using semiconductors diodes.
Material And Methods:
The performance of the two in vivo dosimetry methods was assessed. The primary endpoint was the dose deviation between the reconstructed dose at the prescription point and the measured dose using the ionization chamber in phantoms or the calculated predictive dose by the treatment planning system with patients. The deviation threshold was set to ±5%. In total, 107 patients were prospectively included and treated with 3D-conformal radiotherapy (3D-CRT) or intensity-modulated radiotherapy (IMRT) techniques for tumours of the brain, chest and head and neck.
Results:
The dosimetric accuracy of EPIgray(®) in phantom were comparable to diodes in terms of repeatability (0.11%), reproducibility (0.29-0.51%) with a mean dose deviation of 0.17% (SD: 1.11). The rates of radiotherapy sessions out of the tolerance for the brain (3D-CRT and IMRT), thorax (3D-CRT) and the head and neck (IMRT) were respectively 0%, 9.6% and 5.3% with a mean dose deviation ranging between 0.49% and 1.53%. The mean of dose deviation between three consecutive sessions with EPIgray(®) validates 99.1% of treatments.
Conclusion:
The performance of EPIgray(®) in in vivo dosimetry is consistent with the recommendations of the European Society for Radiotherapy and Oncology (ESTRO) and equivalent to semiconductor diodes for 3D-CRT. It also allows adequate control for IMRT, which is technically difficult to perform with the diodes.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Measurement of Bioavailability: Pharmacodynamic Methods