Related Experiment Video
Updated: Apr 30, 2026

Experimental Protocol for Detecting Cyanobacteria in Liquid and Solid Samples with an Antibody Microarray Chip
Published on: February 7, 2017
Development of a real-time capacitive biosensor for cyclic cyanotoxic peptides based on Adda-specific antibodies
Lesedi Lebogang1, Martin Hedström1, Bo Mattiasson2
1Department of Biotechnology, Lund University, Box 124, 22100 Lund, Sweden.
Abstract:
The harmful effects of cyanotoxins in surface waters have led to increasing demands for accurate early warning methods. This study proposes a capacitive immunosensor for broad-spectrum detection of the group of toxic cyclic peptides called microcystins (∼80 congeners) at very low concentration levels. The novel analytical platform offers significant advances compared to the existing methods. Monoclonal antibodies (mAbs, clone AD4G2) that recognize a common element of microcystins were used to construct the biosensing layer. Initially, a stable insulating anchor layer for the mAbs was made by electropolymerization of tyramine onto a gold electrode surface, with subsequent incorporation of gold nanoparticles (AuNPs) on the glutaraldehyde (5%) activated polytyramine surface. The biosensor responded linearly to microcystin concentrations from 1×10(-13)M to 1×10(-10)M MC-LR standard with a limit of detection of 2.1×10(-14)M. The stability of the biosensor was evaluated by repeated measurements of the antigen and by determining the capacitance change relative to the original response, which decreased below 90% after the 30th cycle.