A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
Validity and reliability of a field technique for sweat Na+ and K+ analysis during exercise in a hot-humid
Lindsay B Baker1, Corey T Ungaro, Kelly A Barnes
1Gatorade Sports Science Institute, Barrington, Illinois, USA.
Abstract:
Abstract This study compared a field versus reference laboratory technique for extracting (syringe vs. centrifuge) and analyzing sweat [Na(+)] and [K(+)] (compact Horiba B-722 and B-731, HORIBA vs. ion chromatography, HPLC) collected with regional absorbent patches during exercise in a hot-humid environment. Sweat samples were collected from seven anatomical sites on 30 athletes during 1-h cycling in a heat chamber (33°C, 67% rh). Ten minutes into exercise, skin was cleaned/dried and two sweat patches were applied per anatomical site. After removal, one patch per site was centrifuged and sweat was analyzed with HORIBA in the heat chamber (CENTRIFUGE HORIBA) versus HPLC (CENTRIFUGE HPLC). Sweat from the second patch per site was extracted using a 5-mL syringe and analyzed with HORIBA in the heat chamber (SYRINGE HORIBA) versus HPLC (SYRINGE HPLC). CENTRIFUGE HORIBA, SYRINGE HPLC, and SYRINGE HORIBA were highly related to CENTRIFUGE HPLC ([Na(+)]: ICC = 0.96, 0.94, and 0.93, respectively; [K(+)]: ICC = 0.87, 0.92, and 0.84, respectively), while mean differences from CENTRIFUGE HPLC were small but usually significant ([Na(+)]: 4.7 ± 7.9 mEql/L, -2.5 ± 9.3 mEq/L, 4.0 ± 10.9 mEq/L (all P < 0.001), respectively; [K(+)]: 0.44 ± 0.52 mEq/L (P < 0.001), 0.01 ± 0.49 mEq/L (P = 0.77), 0.50 ± 0.48 mEq/L (P < 0.001), respectively). On the basis of typical error of the measurement results, sweat [Na(+)] and [K(+)] obtained with SYRINGE HORIBA falls within ±15.4 mEq/L and ±0.68 mEq/L, respectively, of CENTRIFUGE HPLC 95% of the time. The field (SYRINGE HORIBA) method of extracting and analyzing sweat from regional absorbent patches may be useful in obtaining sweat [Na(+)] when rapid estimates in a hot-humid field setting are needed.
Related Concept Videos
Accessory Structures of the Skin: Sweat Glands
Sweat glands are classified as merocrine glands; that is, the secretions are excreted by exocytosis through a duct without affecting the cells of the gland. There...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes