Direct air activation measurements at a 15-MV medical linear accelerator
M K Saeed1, B Poppe2, H W Fischer3
1Physics Department, University of Bremen, Bremen 28359, Germany mohamedrick@gmail.com.
Abstract:
Direct radiometric determination of (14)N (γ, n) (13)N air activation was achieved at a 15-MV medical linear accelerator operating in a high-energy photon mode. (13)N was identified by irradiating a gas-tight Marinelli beaker filled with nitrogen gas and later observing the 10-min half-life of the 511-keV positron-electron annihilation line using high-resolution gamma spectroscopy. Quantitative evaluation of the spectral signal yielded a (13)N production rate of 836.8 ± 32 Bq Gy(-1) in air per 40 × 40 cm(2) field cross section at 100 cm source-surface distance.