Related Experiment Video
Updated: Apr 30, 2026

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Radiation Dosimetry of (18)F-FPEB in Humans
Robert M Kessler1, John Seibyl2, Ronald L Cowan3
1Vanderbilt University School of Medicine, Nashville, Tennessee rkessler@uabmc.edu.
Unlabelled:
(18)F-3-fluoro-5-[(pyridin-3-yl)ethynyl]benzonitrile ((18)F-FPEB) is a potent and specific radioligand for the metabotropic glutamate receptor subtype 5 (mGluR5). Before undertaking clinical research studies with (18)F-FPEB, we performed studies of human radiation dosimetry.
Methods:
Serial whole-body scans were obtained in 9 healthy human subjects (5 men, 4 women) for 190-440 min after the intravenous administration of (18)F-FPEB. Radiation doses were estimated using the OLINDA/EXM software.
Results:
Peak organ doses were to the urinary bladder wall, 0.258 mGy/MBq (0.955 rad/mCi), and gallbladder wall, 0.193 mGy/MBq (0.716 rad/mCi). The effective dose was 0.025 mSv/MBq (0.0922 rem/mCi). The doses to the red marrow and spleen were 0.00797 mGy/MBq (0.0295 rad/mCi) and 0.00709 mGy/MBq (0.0262 rad/mCi), respectively. Reducing the urinary voiding interval to 60 or 90 min lowered the urinary bladder wall dose to 0.0885 mGy/MBq (0.327 rad/mCi) or 0.128 mGy/MBq (0.473 rad/mCi), respectively, and the effective dose to 0.0149 mSv/MBq (0.0551 rem/mCi) or 0.0171 mSv/MBq (0.0634 rem/mCi), respectively.
Conclusion:
Urinary voiding should be performed during (18)F-FPEB studies to minimize radiation exposure to research subjects.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
11:03Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases
Published on: May 29, 2017
Related Concept Videos
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...