Summary
Using a magnet protocol for perioperative management of implantable cardioverter-defibrillators (ICDs) during electrocautery procedures significantly reduces ICD off-time and provider handoffs compared to programming the ICD off.
Area of Science:
- Cardiology
- Medical Devices
- Surgical Procedures
Background:
- Limited data exists for perioperative implantable cardioverter-defibrillator (ICD) management during electrocautery.
- Electrocautery use in patients with ICDs requires careful management to prevent device malfunction.
Purpose of the Study:
- To compare the efficacy of a magnet protocol versus programming ICDs off for perioperative management during electrocautery.
- To evaluate ICD off-time, caregiver handoffs, and electromagnetic interference (EMI).
Main Methods:
- A multicenter randomized controlled trial involving 80 patients with pectoral ICDs undergoing surgery with monopolar electrocautery.
- Patients were randomized to an "Off" group (ICD therapy programmed off) or a "Magnet" group (ICD therapy suspended with a magnet).
- A registry tracked ICD patients with procedures within 6 inches of the ICD.
Main Results:
- The "Magnet" group experienced significantly less "excess" ICD off-time (28 minutes) compared to the "Off" group (115 minutes).
- The "Off" group had more caregiver handoffs (6.6 vs 5.5).
- No electromagnetic interference (EMI), arrhythmic events, or device resets were observed in either group.
Conclusions:
- A magnet protocol simplifies perioperative ICD management for electrocautery procedures over 6 inches from the ICD.