Estimating energy expenditure using heat flux measured at a single body site
Kate Lyden1, Tracy Swibas, Victoria Catenacci
11Division of Endocrinology, Metabolism, and Diabetes, University of Colorado Anschutz Medical Campus, Aurora, CO; 2Division of Geriatric Medicine, University of Colorado Anschutz Medical Campus, Aurora, CO; 3School of Public Health, University of Colorado Anschutz Medical Campus, Aurora, CO; and 4Necessity Consulting, Mars, PA.
Introduction:
The Personal Calorie Monitor (PCM) is a portable direct calorimeter that estimates energy expenditure (EE) from measured heat flux (i.e., the sum of conductive, convective, radiative, and evaporative heat).
Purpose:
The primary aim of this study was to compare EE estimated from measures of heat flux with those measured using indirect calorimetry in a thermoneutral environment (26°C). A secondary aim was to determine whether exposure to ambient temperature below thermoneutral condition (19°C) influences the accuracy of the PCM.
Methods:
Thirty-four adults (mean ± SD: age, 28 ± 5 yr; body mass index, 22.9 ± 2.6 kg · m(-2)) were studied for 5 h in a whole-room indirect calorimeter (IC) in thermoneutral and cool conditions. Participants wore the PCM on their upper arm and completed two 20-min treadmill walking bouts (0% grade, 3 mph). The remaining time was spent sedentary (e.g., watching television, using a computer).
Results:
In thermoneutral conditions, EE values (mean (95% confidence interval)) measured by IC and PCM were 560.0 (526.5-593.5) and 623.3 (535.5-711.1) kcal, respectively. In cool conditions, EE values measured by IC and PCM were 572.5 (540.9-604.0) and 745.5 (668.1-822.8) kcal, respectively. Under thermoneutral conditions, mean PCM minute-by-minute EE tracked closely with IC, resulting in a small nonsignificant bias (63 kcal (-5.8 to 132.4)). During cool conditions, mean PCM minute-by-minute EE did not track IC, resulting in a large bias (173.0 kcal (93.9-252.1)) (P <; 0.001).
Conclusions:
This study demonstrated the validity of using measured heat flux to estimate EE. However, accuracy may be impaired in cool conditions possibly because of excess heat loss from the exposed limbs.
More Related Videos
Related Concept Videos
Quantifying Heat
Body Temperature
Body Temperature
The average body temperature is approximately 37°C (98.6°F) and typically ranges from 36.1–37.2°C...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Absorption of Radiation
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...