Comparison study of upper arm and forearm non-invasive blood pressures in adult Emergency Department patients
Related Experiment Video
Updated: Apr 30, 2026

Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
Karen Schimanski1, Andrew Jull2, Nancy Mitchell3
1Emergency Department, Auckland City Hospital, Private Bag 92024, Auckland, New Zealand.
Background:
Forearm blood pressures have been suggested as an alternative site to measure blood pressures when the upper arm is unavailable. However there is little evidence utilising clinical populations to support this substitution.
Objectives:
To determine agreement between blood pressures measured in the left upper arm and forearm using a singular oscillometric non-invasive device in adult Emergency Department patients. The secondary objective was to explore the relationship of blood pressure differences with age, sex, ethnicity, smoking history and obesity.
Design:
Single centre comparison study.
Setting:
Adult Emergency Department, Tertiary Trauma Centre.
Participants:
Forty-four participants who met inclusion/exclusion criteria selected sequentially from the Emergency Department arrival board.
Methods:
A random assignment of order of measurement for left upper arm and forearm blood pressures was utilised. Participants were eligible if they were aged 18 years or older, had been assigned an Australasian Triage Scale code of 2, 3, 4, or 5, were able to consent, and able to have blood pressures measured on their left arm whilst lying at a 45° angle. The Bland-Altman method of statistical analysis was used, with the level of agreement for clinical acceptability for the systolic, diastolic and mean arterial pressure defined as ±10 mmHg.
Results:
The forearm measure overestimated systolic (mean difference 2.2 mmHg, 95% limits of agreement ±19 mmHg), diastolic (mean difference 3.4 mmHg, 95% limits of agreement ±14.4 mmHg), and mean arterial pressures (mean difference 4.1 mmHg, 95% limits of agreement ±13.7 mmHg). The systolic measure was not significantly different from zero. Evidence of better agreement was found with upper arm/forearm systolic measures below 140 mmHg compared to systolic measures above 140 mmHg using the Levene's test (p=0.002, F-statistic=11.09). Blood pressure disparity was not associated with participant characteristics.
Conclusions:
Forearm measures cannot routinely replace upper arm measures for blood pressure measurement. If the clinical picture requires use of forearm blood pressure, the potential variance from an upper arm measure is ±19 mmHg for systolic pressure, although the variability may be close to ±10 mmHg if the systolic blood pressure is below 140 mmHg.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...