Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Gas Chromatography–Mass Spectrometry (GC–MS)01:14

Gas Chromatography–Mass Spectrometry (GC–MS)

6.0K
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
6.0K
Gas Chromatography: Introduction01:13

Gas Chromatography: Introduction

4.9K
Gas chromatography (GC) is a technique for separating and analyzing volatile compounds in a sample. Its primary purpose is to identify and quantify components in complex mixtures, making it essential in fields such as environmental analysis, pharmaceuticals, and petrochemicals. GC is also called vapor-phase chromatography (VPC) or gas-liquid partition chromatography (GLPC).
In GC,  a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
4.9K
Gas Chromatography: Types of Columns and Stationary Phases01:17

Gas Chromatography: Types of Columns and Stationary Phases

4.1K
Gas chromatography (GC) relies on stationary phases to separate and analyze components in a sample. There are two main types of stationary phases: liquid and solid. Liquid stationary phases are non-volatile, thermally stable, and chemically inert liquids coated onto the column. Solid stationary phases are particles of adsorbent material, such as silica gel or molecular sieves.
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
4.1K
Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

2.6K
Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
2.6K
Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

1.5K
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
1.5K
Gas Chromatography: Sample Injection Systems01:08

Gas Chromatography: Sample Injection Systems

2.8K
In gas chromatography, the sample is introduced as a vapor plug into the carrier gas stream for high efficiency and resolution. A microsyringe injects the sample solution into a heated sample port, vaporizing it and mixing it with the carrier gas. This process is important to ensure the sample is properly prepared for analysis. Thermally sensitive samples can be injected directly into the column and volatilized by slowly increasing the column temperature.
Two primary injection methods are used...
2.8K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comprehensive characterization of European house dust contaminants: Concentrations and profiles, geographical variability, and implications for chemical regulation and health risk.

The Science of the total environment·2024
Same author

Concentrations of potentially endocrine disrupting chemicals in car cabin air and dust - Effect of temperature and ventilation.

The Science of the total environment·2024
Same author

Immunological biomarkers in patients with radiographic axial spondyloarthritis, an exploratory longitudinal Swedish study.

Modern rheumatology·2024
Same author

Pesticide screening of surface water and soil along the Mekong River in Cambodia.

The Science of the total environment·2023
Same author

What is in the fish? Collaborative trial in suspect and non-target screening of organic micropollutants using LC- and GC-HRMS.

Environment international·2023
Same author

Temporal trends of lipophilic organic contaminants in blue mussel (1994-2017) and eelpout (1994-2017) from the southern Baltic Sea.

The Science of the total environment·2023

Related Experiment Video

Updated: Apr 30, 2026

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

8.3K

Multivariate data analysis to characterize gas chromatography columns for dioxin analysis.

Lan Do1, Paul Geladi2, Peter Haglund1

  • 1Department of Chemistry, Umea University, SE-901 87 Umeå, Sweden.

Journal of Chromatography. A
|May 13, 2014
PubMed
Summary
This summary is machine-generated.

This study classified 22 gas chromatography (GC) columns into four selectivity groups for analyzing polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs). Physicochemical and structural properties were correlated with retention times to optimize GC methods for PCDD/F profiling.

Keywords:
Column selectivity evaluationDioxinsGCGC×GCMolecular descriptorsMultivariate data analysis

More Related Videos

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.0K
Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.8K

Related Experiment Videos

Last Updated: Apr 30, 2026

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

8.3K
Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.0K
Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.8K

Area of Science:

  • Analytical Chemistry
  • Environmental Chemistry
  • Chromatography

Background:

  • Polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) are persistent organic pollutants requiring robust analytical methods for detection and profiling.
  • Gas chromatography (GC) is a primary technique for separating complex mixtures of PCDD/Fs, but column selectivity significantly impacts resolution.
  • Understanding the relationship between stationary phase properties and PCDD/F retention is crucial for developing advanced separation techniques like comprehensive two-dimensional GC (GC×GC).

Purpose of the Study:

  • To evaluate the selectivity of 22 different GC columns for the separation of tetra- to hexa-chlorinated PCDD/Fs.
  • To identify key physicochemical and structural descriptors that influence solute-stationary phase interactions.
  • To determine optimal orthogonal column combinations for enhanced PCDD/F separation in GC×GC.

Main Methods:

  • Principal Component Analysis (PCA) was employed to group 22 GC columns based on their selectivity for 128 tetra- to hexa-CDD/F congeners.
  • Partial Least Squares (PLS) regression was used to correlate retention times with molecular descriptors (e.g., size, solubility, charge distribution, reactivity).
  • Linear regression analysis was applied to identify orthogonal column combinations for GC×GC applications.

Main Results:

  • 21 of the 22 GC columns were classified into four distinct selectivity groups based on their stationary phase characteristics (non-polar, ionic liquid, phenyl/cyanopropyl, shape-selective).
  • Molecular size, solubility, charge distribution, and reactivity were identified as the most influential physicochemical parameters affecting PCDD/F retention.
  • Specific column combinations exhibiting orthogonality were identified, suitable for improving PCDD/F separation in GC×GC.

Conclusions:

  • GC column selectivity for PCDD/Fs can be effectively categorized using PCA, revealing distinct groups based on stationary phase chemistry.
  • Physicochemical properties related to molecular size, polarity, and solubility are critical drivers of PCDD/F separation on different GC columns.
  • The identified orthogonal column sets provide a basis for developing more powerful GC×GC methods for comprehensive PCDD/F analysis and profiling.