Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Maximizing MR signal for 2D UTE slice selection in the presence of rapid transverse relaxation
Michael Carl1, Jing-Tzyh Alan Chiang2, Jiang Du3
1GE Healthcare, University of California, San Francisco.
Abstract:
Ultrashort TE (UTE) sequences allow direct visualization of tissues with very short T2 relaxation times, such as tendons, ligaments, menisci, and cortical bone. In this work, theoretical calculations, simulations, and phantom studies, as well as in vivo imaging were performed to maximize signal-to-noise ratio (SNR) for slice selective RF excitation for 2D UTE sequences. The theoretical calculations and simulations were based on the Bloch equations, which lead to analytic expressions for the optimal RF pulse duration and amplitude to maximize magnetic resonance signal in the presence of rapid transverse relaxation. In steady state, it was found that the maximum signal amplitude was not obtained at the classical Ernst angle, but at an either lower or higher flip angle, depending on whether the RF pulse duration or amplitude was varied, respectively.
Related Concept Videos
Atomic Nuclei: Types of Nuclear Relaxation
In spin–lattice or longitudinal relaxation, the excited spins exchange energy with the surrounding lattice as they return to the lower energy level. Among several mechanisms that contribute to spin–lattice relaxation, magnetic dipolar interactions are significant. Here, the excited nucleus transfers...