Utilization of assay performance characteristics to estimate hemoglobin A1c result reliability
Alison Woodworth1, Nichole Korpi-Steiner2, James J Miller3
1Vanderbilt University, Nashville, TN;
Clinical Chemistry
|May 13, 2014
Summary
Assessing hemoglobin A1c (HbA1c) test performance reveals significant variation in reporting unreliable results across different instruments. Robust quality control (QC) is crucial for accurate diabetes management, as assay bias greatly impacts reliability.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Diabetes Diagnostics
Background:
- Hemoglobin A1c (HbA1c) testing accuracy is vital for diabetes management.
- Allowable total error (TE(a)) goals necessitate minimal assay imprecision and bias.
- Effective quality control (QC) programs are essential for reliable HbA1c results.
Purpose of the Study:
- To compare the influence of TE(a) goals, QC practices, and assay performance on reporting unreliable HbA1c results.
- To evaluate six different HbA1c instruments across four academic medical centers.
- To determine the risk of unreliable results based on analytical performance.
Main Methods:
- Applied CLSI protocols EP-5 and EP-9 to assess imprecision and bias.
- Investigated six HbA1c platforms: Bio-Rad (Variant II Turbo, Variant II), Tosoh (G8), Sebia (Capillarys 2 Flex Piercing), Roche (COBAS Integra 800), and Siemens (DCA Vantage).
- Calculated patient-weighted sigma (σ) values and risk of unreliable results at 5%, 6%, and 7% TE(a) specifications.
Main Results:
- Observed a wide range of patient-weighted σ values, spanning 0.5 orders of magnitude at a 6% TE(a).
- While imprecision was consistently low (<3%), assay bias significantly affected σ values.
- Analytical performance alone resulted in nearly a 500-fold variation in estimates for reporting unreliable results.
Conclusions:
- Significant differences exist in the probability of reporting unreliable HbA1c results among NGSP-certified platforms.
- At a 6% TE(a), most platforms require maximum QC implementation, except for the Capillarys 2 Flex Piercing.
- Laboratory-specific risk assessments can guide QC strategies to mitigate unreliable HbA1c results.