Related Experiment Video
Updated: Apr 30, 2026

A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Utilization of assay performance characteristics to estimate hemoglobin A1c result reliability
Alison Woodworth1, Nichole Korpi-Steiner2, James J Miller3
1Vanderbilt University, Nashville, TN;
Background:
Allowable total error (TE(a)) goals for hemoglobin (Hb) A(1c) require minimal assay imprecision and bias and implementation of a robust QC monitoring program. Here, we compare the combined influence on the risk of reporting unreliable results of TE(a) goals, a routine QC practice, and assay performance characteristics of 6 Hb A(1c) instruments across 4 academic medical centers.
Methods:
The CLSI protocols EP-5 and EP-9 were applied to investigate Hb A(1c) result imprecision and bias on the Variant II Turbo and Variant II (Bio-Rad), G8 (Tosoh), Capillarys 2 Flex Piercing (Sebia), COBAS Integra 800 (Roche), and DCA Vantage (Siemens). Patient-weighted σ values and the risk of reporting unreliable Hb A(1c) results were determined for each assay at TE(a) specifications of 5%, 6%, and 7%.
Results:
A large range of patient-weighted σ values spanning 0.5 orders of magnitude at a 6% TE(a) was observed. Although imprecision for all instruments was <3%, bias impacted the majority of the σ changes observed. Estimates for reporting unreliable results varied almost 500-fold based on analytical performance alone.
Conclusions:
Considerable differences in the probability of reporting unreliable Hb A(1c) results between different NGSP (formerly the National Glycohemoglobin Standardization Program)-certified platforms were observed. At a 6% TE(a), our study indicates all but the Capillarys 2 Flex Piercing requires that the maximum affordable QC be run. Risk estimates for individual laboratories' Hb A(1c) methods can be used to assess QC practices and residual risk of an unreliable Hb A(1c) result.

