Microfluidic chip for plasma separation from undiluted human whole blood samples using low voltage contactless
Chia-Chern Chen1, Po-Hsiu Lin, Chen-Kuei Chung
1Department of Family Medicine, St Martin de Porres Hospital, Chiayi City, Taiwan, Republic of China.
Lab on a Chip
|May 13, 2014
Abstract:
A plasma separating biochip is demonstrated using a capillary-driven contactless dielectrophoresis method with low voltage (~1 V) and high frequency induced electrostatics between red blood cells. The polarized red blood cells were aggregated and separated from plasma with a 69.8% volume separation and an 89.4% removal rate of red blood cells.