Related Experiment Video
Updated: Apr 29, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Fatigue properties of the hoffmann-vidal external fixation apparatus
Abstract:
Fatigue tests of the standard HoffmannVidal (H-V) external fixator were performed. After 2,500,000 cycles, fractures occurred in the post attached to the pin bracket. The screws on the pin fixation bracket tend to become loose after cyclic loading test, and periodic tightening of all the fixation screws on a H-V apparatus is recommended under in vivo application. In repeated tightening of individual joints, screws on the pin brackets and adjustable connecting rods fractured. Deformation of the retaining set screws in the universal ball joints caused joint stiffness reduction. Severe abrasive wear of the joint surfaces was observed after repeated tightening. It is concluded that this apparatus may be safely used for four to five six-month clinical applications, provided that the critical parts are replaced between sequential applications. Suggestions for clinical management and recommendations for component modifications are made.
More Related Videos
06:58Method and Instrumented Fixture for Femoral Fracture Testing in a Sideways Fall-on-the-Hip Position
Published on: August 17, 2017
09:26An Efficient and Reproducible Protocol for Distraction Osteogenesis in a Rat Model Leading to a Functional Regenerated Femur
Published on: October 23, 2017
Related Concept Videos
Fatigue
Stability of structures
Fatigue Strength of Concrete
Temperature Dependent Deformation
Impact Loading
In cases of elastic deformation,...
Applications of Stress
The...