Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD
Luuk Debie1, Bernard Broers1, Jurren van Opstal1
1Department of Cardiology, University Hospital Maastricht, 6202 AZ Maastricht, The Netherlands.
Abstract:
We report atrio-ventricular cross talk in a patient with a dual chamber ICD leading to ventricular pacing inhibition. This was observed in an ICD without the ventricular safety pace option, which normally is a sufficient protection against this phenomenon. Cross talk could be prevented effectively by reprogramming ventricular sensitivity to a less sensitive setting.