Comparison of high-volume air sampling equipment for viral aerosol sampling during emergency response
Casey Cooper1, Jeremy Slagley2, James Lohaus3
119th Aerospace Medicine Squadron, Little Rock Air Force Base, Arkansas.
Objective:
This study compared the performance of two high-volume bioaerosol air samplers for viable virus to an accepted standard low-volume sampler. In typical bioaerosol emergency response scenarios, highvolume sampling is essential for the low infective concentrations and large air volumes involved.
Design:
Two high-volume air samplers (XMX/2LMIL and DFU-1000) were evaluated alongside a lowvolume sample (BioSampler). Low and high concentrations (9.3-93.2 agent containing particles per liter of air [ACPLA]) of male-specific coliphage 2 (MS2) virus were released into a 12 m3 aerosol test chamber and collected using the air samplers. The collection media from the samplers were then processed and viable virus was assessed via plaque assay.
Setting:
Aerosol test chamber.
Subjects, Participants:
None.
Interventions:
Collection media and flow rate were modified for the XMX/2L-MIL sampler for viable analysis.
Main Outcome Measures:
Concentration estimates in units of plaque forming units per liter of air (PFU/liter) assessed by the samplers as compared to the levels inside the chamber as evaluated with a slit to agar plate in units of ACPLA. Comparison was made via one-way analysis of variance.
Results:
Both the XMX/2L-MIL and DFU-1000 achieved collection effectiveness equal to or greater than the low-volume air sampler for the evaluated MS2 concentrations. The XMX/2L-MIL reliably collected quantifiable low concentrations of MS2, but the DFU-1000 was unable to do so.
Conclusions:
For emergency response to suspected bioaerosols, the evaluated high-volume samplers are as effective as the standard low-flow sampler and should be considered in conducting a health risk assessment. If low concentrations are expected, then high-flow samplers using liquid collection are preferred.