Related Experiment Video
Updated: Apr 29, 2026

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Changes in the functional flexion axis of the knee before and after total knee arthroplasty using a navigation system
David A J Wilson1, Janie L Astephen Wilson2, Glen Richardson3
1School of Biomedical Engineering, Dalhousie University, Halifax, Canada; Department of Surgery, Dalhousie University, Halifax, Canada.
Abstract:
Long term satisfaction of patients with total knee arthroplasty (TKA) has lagged behind that of total hip arthroplasty. One possible reason is the failure of the artificial joint to recreate natural kinematics of the knee. This study evaluated the pre and post implant functional flexion axis in the knees of 285 total knee arthroplasty patients using a surgical navigation system. Results showed that post-implant there was less femoral rollback early in flexion on the lateral side of the joint than pre-implant. Designing future generations of knee implants to allow for this motion may give patients a more 'natural' feeling knee and may benefit outcomes.

