Iterative reconstruction technique vs filter back projection: utility for quantitative bronchial assessment on
Hisanobu Koyama1, Yoshiharu Ohno, Mizuho Nishio
1Division of Radiology, Department of Radiology, Kobe University Graduate School of Medicine, 7-5-2 Kusunoki-cho, Chuo-ku, Kobe, Hyogo, 650-0017, Japan, hkoyama@med.kobe-u.ac.jp.
Objectives:
The aim of this study was to evaluate the utility of the iterative reconstruction (IR) technique for quantitative bronchial assessment during low-dose computed tomography (CT) as a substitute for standard-dose CT in patients with/without chronic obstructive pulmonary disease.
Methods:
Fifty patients (mean age, 69.2; mean % predicted FEV1, 79.4) underwent standard-dose CT (150mAs) and low-dose CT (25mAs). Except for tube current, the imaging parameters were identical for both protocols. Standard-dose CT was reconstructed using filtered back-projection (FBP), and low-dose CT was reconstructed using IR and FBP. For quantitative bronchial assessment, the wall area percentage (WA%) of the sub-segmental bronchi and the airway luminal volume percentage (LV%) from the main bronchus to the peripheral bronchi were acquired in each dataset. The correlation and agreement of WA% and LV% between standard-dose CT and both low-dose CTs were statistically evaluated.
Results:
WA% and LV% between standard-dose CT and both low-dose CTs were significant correlated (r > 0.77, p < 0.00001); however, only the LV% agreement between SD-CT and low-dose CT reconstructed with IR was moderate (concordance correlation coefficient = 0.93); the other agreement was poor (concordance correlation coefficient <0.90).
Conclusions:
Quantitative bronchial assessment via low-dose CT has potential as a substitute for standard-dose CT by using IR and airway luminal volumetry techniques.
Key Points:
• Quantitative bronchial assessment of COPD using low-dose CT is possible. • Airway luminal volumetry with iterative reconstruction is insusceptible to dose reduction. • Filtered back-projection is susceptible to the effect of dose reduction. • Wall area percentage assessment is easily influenced by dose reduction.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
05:07Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...