Reliable quantification of bisphenol A and its chlorinated derivatives in human urine using UPLC-MS/MS method
N Venisse1, C Grignon1, B Brunet1
1University of Poitiers, CNRS-UMR 7285 IC2MP, School of Medicine and Pharmacy (Department of Analytical Chemistry, Pharmaceutics and Epidemiology), 6 rue de la Milétrie, 86034 Poitiers Cedex, France; University Hospital of Poitiers, Biology-Pharmacy-Public Health Pole, 2 rue de la Milétrie, 86021 Poitiers Cedex, France.
Abstract:
Bisphenol A (BPA), a widespread man-made chemical classified as an endocrine disruptor, is increasingly considered as a major cause of concern for human health. Chlorine present in drinking water may react with BPA to form chlorinated derivatives (ClxBPA), which have demonstrated a heightened level of estrogenic activity. If many epidemiological studies report that more than 90% of people have detectable BPA levels in their urine, then no such study has been undertaken regarding ClxBPA. The purpose of this work is to propose a highly sensitive and accurate analytical method adapted to large-scale biomonitoring studies aimed at assessing exposure to BPA and ClxBPA through the use of human urine. To achieve this, we have comprehensively validated a method using salting-out assisted liquid/liquid extraction (SALLE) coupled to UPLC-MS/MS and isotope dilution quantification, to measure unconjugated BPA and ClxBPA in human urine according to the accepted guidelines. Deutered BPA as well as deutered 2,2'-DCBPA was used as internal standards. The matrix calibration curve ranged from 0.05 to 1.60 ng mL(-1) and from 0.5 to 16.0 ng mL(-1) for ClxBPA and BPA respectively, and provided good linearity (r²>0.99). This method was precise (the intra- and inter-day coefficients of variation were <20% at three different concentrations: 0.05 ng mL(-1), 0.2 ng mL(-1), 0.8 ng mL(-1) and 0.5 ng mL(-1), 2 ng mL(-1), 8 ng mL(-1) for ClxBPA and BPA, respectively) and accurate (bias ranged from -13% to +12%). The limit of quantification, validated at 0.05 ng mL(-1) and 0.5 ng mL(-1) for ClxBPA and BPA respectively when using 300 µL of urine, was found to be suitable for the concentration existing in real samples. The matrix effect and the BPA cross-contamination were also investigated in this study. The analytical method developed in this study is in accordance with the requirements applicable to biomonitoring of BPA and ClxBPA in human urine.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013