Fluoroscopic investigation of Riata® transvenous defibrillator leads
Gurjit Singh1, Marc K Lahiri, Arfaat Khan
1Department of Cardiac Electrophysiology, Henry Ford Hospital, Detroit, MI, United States. dr_gurjit@yahoo.com.
Background:
Implantable cardioverter-defibrillator leads from Riata® family (St. Jude Medical Inc., Sylmar, CA, USA) have been recently recalled by Food and Drug Administration for concerns of a unique type of "inside-out" insulation failure leading to conductor externalization. The objective of this study was to evaluate the prevalence and predictors of conductor externalization in patients implanted with Riata 8 French (Fr) and 7 Fr leads.
Methods:
Patients implanted with Riata® and Riata ST® who were actively followed up in our institution were scheduled for high resolution 3 view fluoroscopy and device interrogation including high voltage (HV) lead impedance testing. Fluoroscopic images were graded as presence of externalization or no externalization.
Results:
Of the 90 patients who underwent screening fluoroscopy, majority had dual coil leads (62.5%) and median duration from the implant time to screening was 79.5 months. Twenty four (26.7%) patients exhibited evidence of lead externalization with 10 (41.6%) of these showing electrical abnormalities at the time of screening. No externalization was seen in the 7 Fr leads. Pacing thresholds were significantly elevated in the externalized cohort compared to non-externalized group (1.42 ± 1.23 vs. 0.93 ± 0.53; p = 0.01). Time since lead implant and lead diameter emerged as significant predictors of lead externalization on univariate analysis with only lead diameter being significant on multivariate analysis (odds ratio 30.68; 4.95-∞, p = 0.001).
Conclusions:
Prevalence of insulation failure exhibiting as conductor externalization is high (26.7%) among the large diameter 8 Fr Riata® leads with a significant proportion of patients manifesting electrical failure. High resolution 3 view fluoroscopy is a reasonable approach to screen for this unique type of insulation failure.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization IV: Nursing Management