Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Pressure Relationships in Thoracic Cavity01:24

Pressure Relationships in Thoracic Cavity

8.3K
Breathing, otherwise known as pulmonary ventilation, is the process of air movement into and out of the lungs. The main mechanisms propelling pulmonary ventilation are atmospheric pressure (Patm), intra-pulmonary (Ppul ) or intra-alveolar pressure (Palv) within the alveoli, and intrapleural pressure (Pip) within the pleural cavity.
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
8.3K
Measurement of Fluid Pressure01:16

Measurement of Fluid Pressure

2.3K
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
2.3K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Pressure Gauges01:20

Pressure Gauges

5.3K
Most pressure gauges, like those on scuba tanks, are calibrated to read zero at atmospheric pressure. Readings from such gauges are called the gauge pressure, which is the pressure relative to atmospheric pressure. When the pressure inside the tank exceeds atmospheric pressure, the gauge reports a positive value. Some gauges are designed to measure negative pressure. For example, many physics experiments must take place in a vacuum chamber, a rigid chamber from which some of the air is pumped...
5.3K
Sites for measuring blood pressure01:21

Sites for measuring blood pressure

3.9K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.9K
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer02:57

Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer

33.9K
Gas pressure is caused by force exerted by gas molecules colliding with the surfaces of objects. Although the force of each collision is very small, any surface of an appreciable area experiences a large number of collisions in a short time, which can result in high pressure.
33.9K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Cediranib with weekly paclitaxel or olaparib versus weekly paclitaxel for advanced or recurrent endometrial cancer (COPELIA): a multicentre, open-label, randomised, phase 2 trial in the UK.

EClinicalMedicine·2026
Same author

Differential cytokine architecture in patients treated with CART19 versus CART22.

Journal for immunotherapy of cancer·2026
Same author

Life lost due to the COVID-19 pandemic: A model-based cohort analysis of mortality displacement in the registered population of England.

PloS one·2026
Same author

Procedural Techniques, Airway Management, and Specimen Acquisition for Peripheral Lung Transbronchial and Endobronchial Ultrasound-Guided Cryobiopsy: A Modified Delphi Consensus Statement.

Chest·2026
Same author

A randomized controlled trial of cryoprobe versus forceps for transbronchial biopsy (FROSTBITE-2): Study protocol for a multi-center pragmatic clinical trial.

Contemporary clinical trials·2026
Same author

Occurrence of trace elements and their impact on human health around waste dumping sites.

Environmental geochemistry and health·2026

Related Experiment Video

Updated: Apr 29, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
08:25

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship

Published on: January 8, 2019

8.9K

Comparison of pleural pressure measuring instruments.

Hans J Lee1, Lonny Yarmus1, David Kidd1

  • 1The Division Pulmonary/Critical Care, Johns Hopkins University, Baltimore, MD.

Chest
|May 24, 2014
PubMed
Summary

A handheld digital manometer (DM) is a valid tool for measuring pleural pressure (Ppl) during thoracentesis, showing strong correlation with electronic transducers. U-tube manometers were less accurate.

More Related Videos

The Measurement of Unsteady Surface Pressure Using a Remote Microphone Probe
08:53

The Measurement of Unsteady Surface Pressure Using a Remote Microphone Probe

Published on: December 3, 2016

6.7K
Measurement of the Pressure-volume Curve in Mouse Lungs
09:49

Measurement of the Pressure-volume Curve in Mouse Lungs

Published on: January 27, 2015

18.5K

Related Experiment Videos

Last Updated: Apr 29, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
08:25

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship

Published on: January 8, 2019

8.9K
The Measurement of Unsteady Surface Pressure Using a Remote Microphone Probe
08:53

The Measurement of Unsteady Surface Pressure Using a Remote Microphone Probe

Published on: December 3, 2016

6.7K
Measurement of the Pressure-volume Curve in Mouse Lungs
09:49

Measurement of the Pressure-volume Curve in Mouse Lungs

Published on: January 27, 2015

18.5K

Area of Science:

  • Pulmonology
  • Medical Devices
  • Diagnostic Procedures

Background:

  • Accurate pleural pressure (Ppl) measurement is crucial during thoracentesis.
  • Traditional methods like U-tube (UT) manometers may have limitations.
  • Electronic transducers (ET) are considered a reference standard.

Purpose of the Study:

  • To compare the accuracy of a handheld digital manometer (DM) against a U-tube (UT) manometer and an electronic transducer (ET) for Ppl measurement during thoracentesis.

Main Methods:

  • Thirty-three patients undergoing thoracentesis were enrolled.
  • Pleural pressure (Ppl) was measured using a handheld DM, UT manometer, and ET.
  • Measurements were taken at various stages, including after fluid aspiration.

Main Results:

  • A strong linear correlation was found between the DM and ET for elastance (r = 0.9582).
  • The correlation between the UT manometer and ET was poor (r = 0.0448).
  • DM and ET measurements showed low correlation in patients experiencing cough (R2 = 0.104).

Conclusions:

  • The handheld DM is a valid and user-friendly method for measuring Ppl during thoracentesis.