Multicenter experience with extraction of the Riata/Riata ST ICD lead
Melanie Maytin1, Bruce L Wilkoff2, Michael Brunner2
1Brigham and Women's Hospital, Boston, Massachusetts.
Extraction of Riata implantable cardioverter-defibrillator leads, particularly those with externalized cables, can be safely performed by experienced operators. High-volume centers achieve success rates comparable to published data, despite challenges.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Safety
Background:
- The Food and Drug Administration issued a class I recall for Riata and Riata ST implantable cardioverter-defibrillator leads in November 2011.
- Management recommendations for the Riata lead recall remain controversial, with limited data on extraction safety and feasibility.
Purpose of the Study:
- To evaluate the safety and feasibility of extracting Riata and Riata ST implantable cardioverter-defibrillator leads.
- To assess procedural success rates and complication profiles associated with Riata lead extraction.
Main Methods:
- A retrospective study was conducted across 11 centers, analyzing data from patients undergoing extraction of Riata/Riata ST leads.
- Data collected included lead type, implant duration, reasons for extraction, lead integrity findings, and procedural outcomes.
Main Results:
- 577 Riata/Riata ST leads were extracted from 577 patients between 2003 and 2013. Complete procedural success was achieved in 99.1%.
- The primary reasons for extraction were infection (53.0%) and lead malfunction (35.7%). Externalized cables were found in 34.9% of evaluated leads, with longer implant durations associated with externalization.
- Major complications occurred in 0.87% of cases, including perforations and pericardial effusion. Laser sheaths were more frequently used for leads with externalized cables.
Conclusions:
- Extraction of Riata/Riata ST leads can be challenging, especially when externalized cables are present, necessitating specific techniques.
- Experienced operators at high-volume centers can safely extract these leads, achieving complication rates comparable to existing literature.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization