[Correlation between end-tidal carbon dioxide and partial pressure of arterial carbon dioxide in ventilated newborns]
Jin-Xing Feng1, Xiao-Hong Liu, Hui-Jun Huang
1Department of Neonatology, Shenzhen Children's Hospital, Shenzhen 518026, China. szfjx2013@hotmail.com.
Objective:
To study the correlation between end-tidal carbon dioxide (PetCO2) and partial pressure of arterial carbon dioxide (PaCO2) in ventilated newborns.
Methods:
Thirty-one ventilated newborn underwent mainstream PetCO2 monitoring; meanwhile, arterial blood gas analysis was performed. The correlation and consistency between PetCO2 and PaCO2 were assessed.
Results:
A total of 85 end-tidal and arterial CO2 pairs were obtained from 31 ventilated newborns. The mean PetCO2 (41±10 mm Hg) was significantly lower than the corresponding mean PaCO2 (46±11 mm Hg) (P<0.01). There was a significant positive correlation between PetCO2 and PaCO2 (r=0.92, P<0.01). The overall PetCO2 bias was 5.1±4.3 mm Hg (95% limits of consistency, -3.3 to 13.6 mmHg), and 5% (4/85) of the points were beyond the 95%CI. When the oxygenation index (OI) was less than 300 mm Hg (n=48), there was a significant positive correlation between PetCO2 and PaCO2 (r=0.85, P<0.01); the PetCO2 bias was 5.9±4.3 mm Hg (95% limits of consistency, -2.6 to 14.5 mm Hg), and 4.2% (2/48) of the points were beyond the 95%CI. When the OI was more than 300 mm Hg (n=37), there was also a significant positive correlation between PetCO2 and PaCO2 (r=0.91, P<0.01); the PetCO2 bias was 4.1±4.1 mm Hg (95% limits of consistency, -3.9 to 12.1 mm Hg), and 5% (2/37) of the points were beyond the 95%CI.
Conclusions:
There is a good correlation and consistency between PetCO2 and PaCO2 in ventilated newborns.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated....
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Respiratory Regulation of Acid-Base Balance
When carbon dioxide levels increase in the blood, more H+ and HCO3⁻ are...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: