Optimization of a validated stability-indicating RP-LC method for the determination of fulvestrant from polymeric
Mehmet Gumustas1, Ceyda Tuba Sengel-Turk, Canan Hascicek
1Ankara University, Faculty of Pharmacy, Department of Analytical Chemistry, Ankara, Turkey.
Abstract:
Fulvestrant is used for the treatment of hormone receptor-positive metastatic breast cancer in postmenopausal women with disease progression following anti-estrogen therapy. Several reversed-phase columns with variable silica materials, diameters, lengths, etc., were tested for the optimization study. A good chromatographic separation was achieved using a Waters X-Terra RP(18) column (250 × 4.6 mm i.d. × 5 µm) and a mobile phase, consisting of a mixture of acetonitrile-water (65:35; v/v) containing phosphoric acid (0.1%). The separation was carried out 40 °C with detection at 215 nm.The calibration curves were linear over the concentration range between 1.0-300 and 1.0-200 µg/mL for standard solutions and biological media, respectively. The proposed method is accurate and reproducible. Forced degradation studies were also realized. This fully validated method allows the direct determination of fulvestrant in dosage form and biological samples. The average recovery of the added fulvestrant amount in the samples was between 98.22 and 104.03%. The proposed method was also applied for the determination of fulvestrant from the polymeric-based nanoparticle systems. No interference from using polymers and other excipients was observed in in vitro drug release studies. Therefore an incorporation efficiency of fulvestrant-loaded nanoparticle could be determined accurately and specifically.
More Related Videos
07:29HPLC Coupled with Chemical Fingerprinting for Multi-Pattern Recognition for Identifying the Authenticity of Clematidis Armandii Caulis
Published on: November 11, 2022
08:33Radiosynthesis of 1-2-[18F]Fluoroethyl-L-Tryptophan using a One-pot, Two-step Protocol
Published on: September 21, 2021
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Elution Process
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: