Clinical efficacy of electronic apex locators: systematic review
Jorge N R Martins1, Duarte Marques2, António Mata3
1School of Dental Medicine, Lisbon University, Lisbon, Portugal; Inter PG Endodontics, New York University College of Dentistry, New York, New York; Department of Endodontics, Implantology Institute, Lisbon, Portugal.
Introduction:
Apical constriction has been proposed as the most appropriate apical limit for the endodontic working length. Despite being the most used, some limitations are attributed to the radiographic method of working length determination. It lacks precision because it is based on the average position of the apical constriction. The electronic apex locators have been presented as an alternative to the odontometry performed by radiography. These devices detect the transition of the pulp to the periodontal tissue, which is anatomically very close to the apical constriction and may perform with improved accuracy.
Methods:
A systematic review was performed to compare the radiographic and electronic methods. Clinical studies that compared both methods were searched for on 7 electronic databases, a manual search was performed on the bibliography of articles collected on the electronic databases, and the authors were contacted to ask for references of more research not detected on the electronic and manual search.
Results:
Twenty-one articles were selected. The majority were comparative or evaluation studies, and very few clinical studies comparing both methods are available. Several methodological limitations are present on the collected articles and debated in this review.
Conclusions:
Although the available scientific evidence base is short and at considerable risk of bias, it is still possible to conclude that the apical locator reduces the patient radiation exposure and also that the electronic method may perform better on the working length determination. At least one radiographic control should be performed to detect possible errors of the electronic devices.
More Related Videos
04:50Operative Technique and Nuances for the Stereoelectroencephalographic SEEG Methodology Utilizing a Robotic Stereotactic Guidance System
Published on: June 9, 2023
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
Related Concept Videos
Electronic Distance Measuring Instruments
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Types of Global Positioning System Surveys
Field Application of Global Positioning System