[Disinfection and recontamination of rigid endoscopes: improved safety using an immersion quiver system]
C Rohrmeier1, J Strutz, W Schneider-Brachert
1Klinik für Hals-Nasen-Ohren-Heilkunde, Universität Regensburg, Franz-Josef-Strauß-Allee 11, 93053, Regensburg, Deutschland, christian.rohrmeier@ukr.de.
Background:
In Otorhinolaryngology, rigid endoscopes are used daily at a high frequency. There is no consensus for reprocessing these medical instruments. Often immersion disinfection procedures are used. The present study examined the possible risk of recontamination by this disinfection method and investigated the possibility of avoiding this risk by using a new immersion quiver system.
Methods:
Using coloured markers, a possible contact of the endoscope with the top edges of quivers of different diameters during endoscope removal was tested for. In addition, it was evaluated whether Staphylococcus aureus transfer is possible via this route. The same methodology was applied to a new immersion quiver system.
Results:
Whenever removing the rigid endoscopes from the conventional quiver, these touched the top of the quiver, regardless of its diameter. A transfer of Staphylococcus aureus from the quiver to the endoscope via this route could be detected in five out of eight attempts. During endoscope removal from the new immersion quiver system, no contact of the endoscope with the outer quiver occurred in 20 passes. In none of eight trials was a transfer of Staphylococcus aureus from previously contaminated immersion quivers to the endoscope shown; all immersion quivers were sterile after disinfection.
Discussion:
After conventional immersion disinfection, recontamination of rigid endoscopes by a contaminated quiver edge is possible. An immersion quiver system can resolve this risk of recontamination easily, by decontaminating not only the endoscope, but also the immersion quiver (inner quiver) itself in the disinfectant solution.
Insights
Recontamination of rigid endoscopes during reprocessing is a risk with conventional immersion disinfection. A new immersion quiver system effectively prevents endoscope recontamination by ensuring both the endoscope and quiver are disinfected.
Area of Science:
- Otorhinolaryngology
- Medical Device Reprocessing
- Infection Control
Context:
- Rigid endoscopes are frequently used in Otorhinolaryngology.
- Current reprocessing methods, particularly immersion disinfection, lack consensus and may pose recontamination risks.
- The study addresses the need for standardized and safe reprocessing protocols for medical instruments.
Purpose:
- To evaluate the risk of recontamination associated with conventional immersion disinfection of rigid endoscopes.
- To assess the efficacy of a novel immersion quiver system in preventing endoscope recontamination.
- To determine if Staphylococcus aureus transfer occurs from conventional quivers to endoscopes post-disinfection.
Summary:
- Conventional immersion disinfection of rigid endoscopes can lead to recontamination via contact with contaminated quiver edges.
- A new immersion quiver system demonstrated no contact between the endoscope and quiver during removal.
- The novel system successfully prevented Staphylococcus aureus transfer and ensured quiver sterility after disinfection.
Impact:
- Highlights a significant risk in current medical instrument reprocessing.
- Introduces a practical solution to enhance patient safety in Otorhinolaryngology.
- Provides evidence for the adoption of improved disinfection systems to prevent healthcare-associated infections.
More Related Videos
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
14:17The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...