Retrieval of trapped and broken guide wire with immediate rescue off-pump coronary bypass surgery
Devender Singh1, Anshuman Darbari2
1Department of Cardiothoracic Surgery, Kovai Medical Centre and Hospital, Coimbatore, Tamilnadu, India.
Abstract:
The entrapment, fracture and dislodgement of diagnostic or therapeutic devices within the coronary circulation during a procedure are a rare complication occurring in 0.2-0.8% of cases. Despite technological improvements, this complication is still occurring because coronary angioplasty is often undertaken for complex anatomical situations. The complication of device fracture during the intervention procedure occurs due to entrapment, overcoiling and excessive traction of the guide wire. There has been no agreement as to whether and by which technique the immediate removal of the broken fragment of guide wire should be done. Here, we report a case of anterolateral myocardial infarction who underwent primary percutaneous coronary intervention. During the procedure, the guide wire was entrapped within the left anterior descending coronary artery. Despite many attempts, the wire could not be removed and even became fractured at the femoral insertion site; thus, urgent surgical removal of the wire with vessel grafting was done with a successful outcome. This gives a clear message about the importance of the ready availability of surgical backup and, particularly, the necessity for complex percutaneous interventions.
More Related Videos
14:35Post-Myocardial Infarction Heart Failure in Closed-chest Coronary Occlusion/Reperfusion Model in Göttingen Minipigs and Landrace Pigs
Published on: April 17, 2021
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014