Anaesthetic conserving device AnaConDa: dead space effect and significance for lung protective ventilation
L W Sturesson1, M Bodelsson2, B Jonson3
1Section of Anaesthesiology and Intensive Care, Lund University and Skane University Hospital, SE-221 85 Lund, Sweden louise.walther_sturesson@med.lu.se.
Background:
The anaesthetic conserving device AnaConDa (ACD) reflects exhaled anaesthetic agents thereby facilitating the use of inhaled anaesthetic agents outside operating theatres. Expired CO₂ is, however, also reflected causing a dead space effect in excess of the ACD internal volume. CO₂ reflection from the ACD is attenuated by humidity. This study tests the hypothesis that sevoflurane further attenuates reflection of CO₂. An analysis of clinical implications of our findings was performed.
Methods:
Twelve postoperative patients received mechanical ventilation using a conventional heat and moisture exchanger (HME, internal volume 50 ml) and an ACD (100 ml), the latter with or without administration of sevoflurane. The ACD was also studied with a test lung at high sevoflurane concentrations. Reflection of CO₂ and dead space effects were evaluated with the single-breath test for CO2.
Results:
Sevoflurane reduced but did not abolish CO₂ reflection. In patients, the mean dead space effect with 0.8% sevoflurane was 88 ml larger using the ACD compared with the HME (P<0.001), of which 38 ml was due to CO₂ reflection. Our calculations show that with the use of the ACD, normocapnia cannot be achieved with tidal volume <6 ml kg(-1) even when respiratory rate is increased.
Conclusions:
An ACD causes a dead space effect larger than its internal volume due to reflection of CO₂, which is attenuated but not abolished by sevoflurane administration. CO₂ reflection from the ACD limits its use with low tidal volume ventilation, such as with lung protection ventilation strategies.
Clinical Trial Registration:
Clinical Trials NCT01699802.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...