Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control
Daniel J DeSalvo1, Satya Shanmugham1, Trang T Ly1
1Stanford University School of Medicine, Stanford, CA, USA.
Insights
The Bayer Contour® Next (CN) glucose meter showed superior accuracy compared to the HemoCue® (HC) system in children with type 1 diabetes during overnight studies. CN met all accuracy standards, making it ideal for future research.
Area of Science:
- Pediatric Endocrinology
- Medical Device Technology
- Diabetes Management
Background:
- Accurate glucose monitoring is crucial for managing type 1 diabetes, especially in children.
- Overnight closed-loop studies require reliable glucose monitoring systems for safety and data integrity.
- Evaluating the accuracy of different glucose monitoring systems is essential for clinical research.
Purpose of the Study:
- To compare the accuracy of the Bayer Contour® Next (CN) and HemoCue® (HC) glucose monitoring systems.
- To assess system performance against Yellow Springs Instrument (YSI) plasma reference values.
- To determine suitability of these systems for overnight closed-loop studies in pediatric type 1 diabetes patients.
Main Methods:
- Pilot study involving 28 children (10-18 years) with type 1 diabetes in a clinical research center.
- Continuous glucose measurements obtained every 30 minutes overnight.
- Comparison of whole blood glucose readings (CN, HC) with plasma reference values (YSI) using Mean Absolute Relative Difference (MARD) and International Organization for Standardization (ISO) criteria.
Main Results:
- The CN system demonstrated significantly better accuracy than the HC system (MARD 3.13% vs. 10.73%, P < .001).
- The CN system met ISO 2003 and 2013 accuracy standards across all glucose ranges.
- The HC system did not meet ISO accuracy criteria within the study's limited sample size.
Conclusions:
- The Bayer Contour® Next glucose monitoring system exhibits high accuracy and reliability in pediatric type 1 diabetes patients.
- CN is a suitable and recommended device for use in overnight closed-loop studies, even outside of a research setting.
- Further research may be warranted to explore HC system performance under different conditions.
Abstract:
This pilot study evaluated the difference in accuracy between the Bayer Contour® Next (CN) and HemoCue® (HC) glucose monitoring systems in children with type 1 diabetes participating in overnight closed-loop studies. Subjects aged 10-18 years old were admitted to a clinical research center and glucose values were obtained every 30 minutes overnight. Glucose values were measured using whole blood samples for CN and HC readings and results were compared to Yellow Springs Instrument (YSI) reference values obtained with plasma from the same sample. System accuracy was compared using mean absolute relative difference (MARD) and International Organization for Standardization (ISO) accuracy standards. A total of 28 subjects were enrolled in the study. Glucose measurements were evaluated at 457 time points. CN performed better than HC with an average MARD of 3.13% compared to 10.73% for HC (P < .001). With a limited sample size, CN met ISO criteria (2003 and 2013) at all glucose ranges while HC did not. CN performed very well, and would make an excellent meter for future closed-loop studies outside of a research center.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Feedback Loops
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Hypoglycemia and Glucagon