Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control

Daniel J DeSalvo1, Satya Shanmugham1, Trang T Ly1

  • 1Stanford University School of Medicine, Stanford, CA, USA.

Insights

The Bayer Contour® Next (CN) glucose meter showed superior accuracy compared to the HemoCue® (HC) system in children with type 1 diabetes during overnight studies. CN met all accuracy standards, making it ideal for future research.

Area of Science:

  • Pediatric Endocrinology
  • Medical Device Technology
  • Diabetes Management

Background:

  • Accurate glucose monitoring is crucial for managing type 1 diabetes, especially in children.
  • Overnight closed-loop studies require reliable glucose monitoring systems for safety and data integrity.
  • Evaluating the accuracy of different glucose monitoring systems is essential for clinical research.

Purpose of the Study:

  • To compare the accuracy of the Bayer Contour® Next (CN) and HemoCue® (HC) glucose monitoring systems.
  • To assess system performance against Yellow Springs Instrument (YSI) plasma reference values.
  • To determine suitability of these systems for overnight closed-loop studies in pediatric type 1 diabetes patients.

Main Methods:

  • Pilot study involving 28 children (10-18 years) with type 1 diabetes in a clinical research center.
  • Continuous glucose measurements obtained every 30 minutes overnight.
  • Comparison of whole blood glucose readings (CN, HC) with plasma reference values (YSI) using Mean Absolute Relative Difference (MARD) and International Organization for Standardization (ISO) criteria.

Main Results:

  • The CN system demonstrated significantly better accuracy than the HC system (MARD 3.13% vs. 10.73%, P < .001).
  • The CN system met ISO 2003 and 2013 accuracy standards across all glucose ranges.
  • The HC system did not meet ISO accuracy criteria within the study's limited sample size.

Conclusions:

  • The Bayer Contour® Next glucose monitoring system exhibits high accuracy and reliability in pediatric type 1 diabetes patients.
  • CN is a suitable and recommended device for use in overnight closed-loop studies, even outside of a research setting.
  • Further research may be warranted to explore HC system performance under different conditions.

Related Concept Videos

Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.6K
Feedback Loops01:01

Feedback Loops

In most cases, excessive hormone production is prevented by negative feedback—a loop that starts with a stimulus inducing the release of a particular substance, like a hormone, to maintain a certain level before triggering a signal that results in a decrease in further release of the hormone.
43.7K
Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.7K
Open and closed-loop control systems01:17

Open and closed-loop control systems

Control systems are foundational elements in automation and engineering. They are broadly categorized into open-loop and closed-loop systems. These classifications hinge on the presence or absence of feedback mechanisms, significantly influencing the system's performance, complexity, and application.
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
1.9K
Glucose Homeostasis: Regulation of Blood Glucose01:02

Glucose Homeostasis: Regulation of Blood Glucose

Carbohydrates consumed through foods are converted into glucose, a crucial energy source for the body. In the prandial state, high blood glucose levels stimulate the secretion of insulin from the pancreas. Insulin inhibits hepatic glucose production and stimulates glucose uptake and metabolism by muscle and adipose tissue. The excess glucose is converted into glycogen and stored in the liver and muscles.
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
4.3K
Hypoglycemia and Glucagon01:15

Hypoglycemia and Glucagon

Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
1.3K