Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?
Sarvenaz Pourjabbar1, Sarabjeet Singh1, Atul Padole1
1Sarvenaz Pourjabbar, Sarabjeet Singh, Atul Padole, Akshay Saini, Michael A Blake, Mannudeep K Kalra, Department of Radiology, Massachusetts General Hospital, Boston, MA 02114, United States.
Body dimensions from CT localizer radiographs can overestimate patient size, leading to underestimation of Size Specific Dose Estimate (SSDE). Accurate dose assessment requires careful consideration of measurement methods for radiation protection.
Area of Science:
- Radiological Physics
- Medical Imaging
- Radiation Dosimetry
Background:
- Accurate radiation dose assessment in computed tomography (CT) is crucial for patient safety.
- Size Specific Dose Estimate (SSDE) aims to provide a more personalized dose metric than CT Dose Index Volume (CTDIvol).
- Body dimensions are key inputs for calculating SSDE.
Purpose of the Study:
- To evaluate the impact of body dimensions derived from localizer radiographs and transverse abdominal CT images on Size Specific Dose Estimate (SSDE).
- To compare the accuracy of patient size measurements from different imaging modalities for dose estimation.
Main Methods:
- Fifty patients undergoing abdominal CT were included.
- Anterior-posterior (AP) and lateral (Lat) diameters were measured from localizer radiographs and transverse CT images.
- Conversion factors were used to calculate SSDE from CTDIvol, considering various diameter measurements and patient centering.
Main Results:
- In 11% of measurements, accurate diameter measurements were not possible due to incomplete body coverage.
- Forty-one patients were off-centered from the gantry iso-center.
- SSDE values showed significant variation compared to CTDIvol, ranging from 32% lower to 74% higher.
Conclusions:
- Localizer radiographs tend to overestimate patient size, leading to an underestimation of SSDE.
- There is a notable discrepancy between SSDE and CTDIvol, highlighting the need for precise body dimension measurements.
- This study underscores the importance of accurate patient sizing for reliable dose estimation in CT.
More Related Videos
09:00Spatial Measurements of Perfusion, Interstitial Fluid Pressure and Liposomes Accumulation in Solid Tumors
Published on: August 18, 2016
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET