Imaging dose in variable pitch body perfusion CT scans: an analysis using TG111 formalism
Marian Axente1, Dimitre Hristov1
1Department of Radiation Oncology, Stanford University Medical Center, Stanford, California 94305-5847.
Purpose:
To investigate the variation of imaging dose with tube potential in variable pitch body CT perfusion (CTp) protocols using the TG111 dosimetric formalism.
Methods:
TG111 recommendations were followed in choosing the phantom, dosimetric equipment, and methodology. Specifically, equilibrium doses (D(eq)) were measured centrally and peripherally in a long PMMA phantom. Reference planar average equilibrium doses were determined for each tube potential, for a reference set of exposure parameters (collimation, pitch, filtration) on a Siemens Definition CT scanner. These reference values were utilized to predict the imaging dose during perfusion scans using interpretations of the TG111 formalism. As a gold reference, the midscan average planar perfusion doses (D(CTp)) were obtained directly from central and peripheral D(eq) measurements for body CTp scans (144 and 271 mm) using variable pitch acquisition. Measurement-based D(CTp) values obtained using a thimble chamber were compared to the TG111-predicted values, and to CTDI(vol) reported at the console.
Results:
Reference planar average equilibrium dose values measured for reference uniform pitch helical scans were consistently higher than console-reported or measured values for CTDI(vol). The measurement-based perfusion dose D(CTp) was predicted accurately by the reported CTDI(vol) for the 144 mm scan. The 271 mm scans delivered systematically larger dose than reported. The TG111-based dose estimates were proven to be conservative, as they were systematically higher than both the measured and the reported imaging doses.
Conclusions:
Upon successful implementation of TG111 formalism, standard imaging dose was measured for a body CTp protocol using the variable pitch helical acquisition. The TG111 formalism is not directly applicable to this type of acquisition. Measurement of dose for all variable pitch protocols is strongly suggested.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:04Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
Published on: April 14, 2011
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Formal Charges
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Lewis Structures and Formal Charges