Simultaneous whole-body PET/MR imaging in comparison to PET/CT in pediatric oncology: initial results
Jürgen F Schäfer1, Sergios Gatidis, Holger Schmidt
1From the Department of Radiology, Diagnostic and Interventional Radiology (J.F.S., S.G., H.S., B.G., C.A.P., C.D.C., N.F.S.), Department of Radiology, Preclinical Imaging and Radiopharmacy, Laboratory for Preclinical Imaging and Imaging Technology of the Werner Siemens Foundation (H.S., I.B.), Department of Radiology, Division of Nuclear Medicine (M.R.), Children's Hospital, Department of General Pediatrics, Hematology and Oncology (M.E.), and Children's Hospital, Department of Pediatric Surgery and Pediatric Urology (J.F.), Eberhard-Karls-University Tübingen, Hoppe-Seyler-Str 3, 72076 Tübingen, Germany; and Department of Empirical Inference, Max Planck Institute for Intelligent Systems, Tübingen, Germany (I.B.).
Purpose:
To compare positron emission tomography (PET)/magnetic resonance (MR) imaging and PET/computed tomography (CT) for lesion detection and interpretation, quantification of fluorine 18 ((18)F) fluorodeoxyglucose (FDG) uptake, and accuracy of MR-based PET attenuation correction in pediatric patients with solid tumors. Materials and Methods This prospective study had local ethics committee and German Federal Institute for Drugs and Medical Devices approval. Written informed consent was obtained from all patients and legal guardians. Twenty whole-body (18)F-FDG PET/CT and (18)F-FDG PET/MR examinations were performed in 18 pediatric patients (median age, 14 years; range, 11-17 years). (18)F-FDG PET/CT and (18)F-FDG PET/MR data were acquired sequentially on the same day for all patients. PET standardized uptake values (SUVs) were quantified with volume of interest measurements in lesions and healthy tissues. MR-based PET attenuation correction was compared with CT-derived attenuation maps (µ-maps). Lesion detection was assessed with separate reading of PET/CT and PET/MR data. Estimates of radiation dose were derived from the applied doses of (18)F-FDG and CT protocol parameters. Descriptive statistical analyses were performed to report correlation coefficients and relative deviations for comparison of SUVs, rates of lesion detection, and percentage reductions in radiation dose.
Results:
PET SUVs showed strong correlations between PET of PET/CT (PETCT) and PET of PET/MR (PETMR) (r > 0.85 for most tissues). Apart from drawbacks of MR-based PET attenuation correction in osseous structures and lungs, similar SUVs were found on PET images corrected with CT-based µ-maps (13.1% deviation of SUVs for bone marrow and <5% deviation for other tissues). Lesion detection rate with PET/MR imaging was equivalent to that with PET/CT (61 areas of focal uptake on PETMR images vs 62 areas on PETCT images). Advantages of PET/MR were observed especially in soft-tissue regions. Furthermore, PET/MR offered significant dose reduction (73%) compared with PET/CT.
Conclusion:
Pediatric oncologic PET/MR is technically feasible, showing satisfactory performance for PET quantification with SUVs similar to those of PET/CT. Compared with PET/CT, PET/MR demonstrates equivalent lesion detection rates while offering markedly reduced radiation exposure. Thus, PET/MR is a promising modality for the clinical work-up of pediatric malignancies. Online supplemental material is available for this article.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...