Sleeping with the fishes: electromagnetic interference causing an inappropriate implantable cardioverter
Hui Min Knight1, Holly Elizabeth Cakebread2, Parag Ravindra Gajendragadkar2
1Department of Medicine, Bedford Hospital, Bedford, UK.
Abstract:
A 60-year-old man with a cardiac defibrillator implanted due to previous ventricular fibrillation arrest and ischaemic cardiomyopathy received a shock while cleaning his fish pond. At the time, his immersed arm was close to a submersed water pump, but the patient was asymptomatic. As a result of the shock he lost consciousness, but collapsed backwards, away from the pond. Interrogation of the device revealed a high-frequency artefact that was sensed by the device and triggered a shock. Device parameters were otherwise normal. Subsequently, the submersed water pump was found to be the source of an external alternating current leak and was identified as the likely cause of the inappropriate shock due to electromagnetic interference (EMI). Awareness of potential sources of EMI along with evaluation of data with a detailed clinical history is warranted in all cases.
More Related Videos
08:43Advanced Cardiac Rhythm Management by Applying Optogenetic Multi-Site Photostimulation in Murine Hearts
Published on: August 26, 2021
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanism of Cardiac Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy V: Interprofessional Care