Robotic calibration of the motional Stark effect diagnostic on Alcator C-Mod
Robert T Mumgaard1, Steven D Scott2, Jinseok Ko3
1MIT Plasma Science and Fusion Center, Cambridge, Massachusetts 02139, USA.
Abstract:
The capability to calibrate diagnostics, such as the Motional Stark Effect (MSE) diagnostic, without using plasma or beam-into-gas discharges will become increasingly important on next step fusion facilities due to machine availability and operational constraints. A robotic calibration system consisting of a motorized three-axis positioning system and a polarization light source capable of generating arbitrary polarization states with a linear polarization angle accuracy of <0.05° has been constructed and has been used to calibrate the MSE diagnostic deployed on Alcator C-Mod. The polarization response of the complex diagnostic is shown to be fully captured using a Fourier expansion of the detector signals in terms of even harmonics of the input polarization angle. The system's high precision robotic control of position and orientation allow it to be used also to calibrate the geometry of the instrument's view. Combined with careful measurements of the narrow bandpass spectral filters, this system fully calibrates the diagnostic without any plasma discharges. The system's high repeatability, flexibility, and speed has been exploited to quantify several systematics in the MSE diagnostic response, providing a more complete understanding of the diagnostic performance.