A Microfluidic Chip for the Versatile Chemical Analysis of Single Cells
Published on: October 15, 2013
Immunological analyses of whole blood via "microfluidic drifting" based flow cytometric chip
Ahmad Ahsan Nawaz1, Ruth Helmus Nissly, Peng Li
1Department of Engineering Science and Mechanics, The Pennsylvania State University, University Park, PA, 16802, USA.
Abstract:
Cost-effective, high-performance diagnostic instruments are vital to providing the society with accessible, affordable, and high-quality healthcare. Here we present an integrated, "microfluidic drifting" based flow cytometry chip as a potential inexpensive, fast, and reliable diagnostic tool. It is capable of analyzing human blood for cell counting and diagnosis of diseases. Our device achieves a throughput of ~3754 events/s. Calibration with Flow-Check calibration beads indicated good congruency with a commercially available benchtop flow cytometer. Moreover, subjection to a stringent 8-peak rainbow calibration particle test demonstrated its ability to perform high-resolution immunological studies with separation resolution of 4.28 between the two dimmest fluorescent populations. Counting accuracy at different polystyrene bead concentrations showed strong correlation (r=0.9991) with hemocytometer results. Finally, reliable quantification of CD4+ cells in healthy human blood via staining with monoclonal antibodies was demonstrated. These results demonstrate the potential of our microfluidic flow cytometry chip as an inexpensive yet high-performance point-of-care device for mobile medicine.