In vitro evaluation of temperature changes in the root canal induced by ultrasonic scalers
U Van der Velden1, T J G Koster, A J Feilzer
1Department of Periodontology, Academic Centre for Dentistry Amsterdam (ACTA), Amsterdam, The Netherlands.
Objective:
To evaluate in vitro the thermal effects induced by four different ultrasonic scalers on the temperature in the root canal during ultrasonic scaling.
Methods:
An extracted lower central incisor provided with a thermocouple in the root canal and a tube, entering the tooth incisally and exiting it apically to simulate an artificial bloodstream, was placed in a model of the lower jaw with soft artificial gingiva. Tested ultrasonic scaler systems included: EMS PM-600, Satelec P-max, Dürr Vector and Dentsply Cavitron. The tooth was scaled with each system at full water supply of 21°C. Furthermore, the amount of water supply was determined to maintain during scaling a constant temperature in the root canal. Finally, thermal changes due to scaling without water were assessed.
Results:
Except for the Vector all scaler systems showed a temperature decrease in the root canal. The Vector with water/polish suspension showed a trend towards an increase in temperature. To maintain a constant temperature in the root canal the Cavitron needed twice the amount of water compared with PM-600 and P-max. Without water, all scaling systems induced a temperature increase.
Conclusion:
For safe ultrasonic scaling, care should be taken that the cooling water has room temperature and that, dependent on the scaler system, the proper amount of water is supplied.