Perforation and bacterial contamination of microscope covers in lumbar spinal decompressive surgery

Georg Osterhoff1, José Spirig, Jürgen Klasen

  • 1Department of Orthopedics, Balgrist Clinic, University of Zurich, Zurich, Switzerland.

Abstract

Insights

Microscope covers used in lumbar spinal decompression surgery showed a low perforation rate (4%) with no link to bacterial contamination. External contamination sources appear more significant than cover integrity issues.

Area of Science:

  • Neurosurgery
  • Surgical Technology
  • Infection Control

Background:

  • Maintaining a sterile surgical field is critical during spinal decompression.
  • Surgical microscopes are essential tools in these procedures, requiring sterile covers.

Purpose of the Study:

  • To assess the integrity of surgical microscope covers.
  • To evaluate bacterial contamination on microscope covers post-lumbar spinal decompression surgery.

Main Methods:

  • Prospective study of 25 lumbar spinal decompressions using surgical microscopes.
  • Microscope covers were tested for perforations using a water leakage test.
  • Bacterial contamination was assessed via swabs taken from covers before and after surgery.

Main Results:

  • One perforation (4%) was detected in 25 microscope covers.
  • Post-operative bacterial contamination was found in 3 (4%) of smears, unrelated to perforations.
  • No association was observed between cover perforation and bacterial contamination.

Conclusions:

  • The incidence of microscope cover perforation during lumbar spinal decompression is low.
  • Cover integrity failure does not appear to be a significant source of bacterial contamination.
  • External contamination sources are likely more influential than microscope cover integrity.