Related Experiment Video
Updated: Apr 28, 2026

Minimally Invasive Murine Laryngoscopy for Close-Up Imaging of Laryngeal Motion During Breathing and Swallowing
Published on: December 1, 2023
Comparing GlideScope Video Laryngoscope and Macintosh Laryngoscope Regarding Hemodynamic Responses During Orotracheal
Ali Reza Pournajafian1, Mohammad Reza Ghodraty1, Seyed Hamid Reza Faiz2
1Department of Anaesthesiology, Firoozgar Hospital, Iran University of Medical Sciences, Tehran, IR Iran.
Background:
To determine if the GlideScope® videolaryngoscope (GVL) could attenuate the hemodynamic responses to orotracheal intubation compared with conventional Macintosh laryngoscope.
Objectives:
The aim of this relatively large randomized trial was to compare the hemodynamic stress responses during laryngoscopy and tracheal intubation using GVL versus MCL amongst healthy adult individuals receiving general anesthesia for elective surgeries.
Patients And Methods:
Ninety five healthy adult patients with American Society of Anesthesiologists physical status class I or II that were scheduled for elective surgery under general anesthesia were randomly allocated to either Macintosh or GlideScope arms. All patients received a standardized protocol of general anesthesia. Hemodynamic changes associated with intubation were recorded before and at 1, 3 and 5 minutes after the intubation. The time taken to perform endotracheal intubation was also noted in both groups.
Results:
Immediately before laryngoscopy (pre-laryngoscopy), the values of all hemodynamic variables did not differ significantly between the two groups (All P values > 0.05). Blood pressures and HR values changed significantly over time within the groups. Time to intubation was significantly longer in the GlideScope (15.9 ± 6.7 seconds) than in the Macintosh group (7.8 ± 3.7 sec) (P< 0.001). However, there were no significant differences between the two groups in hemodynamic responses at all time points.
Conclusions:
The longer intubation time using GVL suggests that the benefit of GVL could become apparent if the time taken for orotracheal intubation could be decreased in GlideScope group.