Published on: December 18, 2016
Flow-induced signal misallocation artifacts in two-point fat-water chemical shift MRI
Mahdi Salmani Rahimi1, James H Holmes, Kang Wang
1Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Purpose:
Two-point fat-water separation methods are increasingly being used for chest and abdominal MRI and have recently been introduced for use in MR angiography of the lower extremities. With these methods, flowing spins can accumulate unintended phase shifts between the echo times. The purpose of this study is to demonstrate that these phase shifts can lead to inaccurate signals in the water and fat images.
Theory And Methods:
In vitro experiments were conducted at 1.5T and 3.0T using a stenosis-mimicking phantom and a computer-controlled pump to image a range of physiologically relevant velocities.
Results:
In the phantom images acquired using bipolar readout gradients, fat-water signal inaccuracies were visible in regions of flow, with increasing severity as the flow rate was increased. Additionally, similar effects were observed in regions of high flow in clinical chest and liver exams. In the phantom images, the effect was eliminated by using a dual-pass method without bipolar readout gradients.
Conclusion:
When using fat-water separation methods with bipolar readout gradients, phase shifts caused by the motion of spins can lead to signal inaccuracies in the fat and water images. These artifacts can be mitigated by using approaches that do not use bipolar readout gradients.
Related Concept Videos
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
Magnetic Resonance Imaging
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
¹H NMR Signal Multiplicity: Splitting Patterns