Optimized Analysis of In Vivo and In Vitro Hepatic Steatosis
Published on: March 11, 2017
Controlled attenuation parameter for evaluating liver steatosis in chronic viral hepatitis
Giovanna Ferraioli1, Carmine Tinelli1, Raffaella Lissandrin1
1Giovanna Ferraioli, Raffaella Lissandrin, Mabel Zicchetti, Gaetano Filice, Carlo Filice, Department of Infectious Diseases, Fondazione IRCCS Policlinico San Matteo, Medical School University of Pavia, 27100 Pavia, Italy.
Aim:
To assess the performance of controlled attenuation parameter (CAP) in patients with chronic viral hepatitis.
Methods:
CAP is a new technique that measures the attenuation in the liver of an ultrasound beam, which is directly related to lipid accumulation. Consecutive patients undergoing liver biopsy for chronic viral hepatitis were studied using the M probe of FibroScan device (Echosens, Paris, France). The device estimates liver steatosis in decibel per meter (dB/m). An expert operator performed all measurements. Steatosis was graded according to Kleiner's classification. Pearson or Spearman rank coefficient was used to test correlation between two study variables. Linear regression was used for multivariate model to assess the association between CAP and other variables. Receiver operating characteristic curve analysis was performed to calculate area under the curve (AUROC) for S0 vs S1-S3 and S0-S1 vs S2-S3.
Results:
115 subjects (85 males and 30 females) were prospectively studied. The mean values of CAP were 227.1 ± 43.1 for S0; 254.6 ± 38.9 for S1; 297.8 ± 49.4 dB/m for S2-S3. In univariate analysis CAP showed a significant correlation with age, body mass index (BMI), degree of steatosis, and cholesterol. Multivariate regression analysis confirmed the correlation with the degree of steatosis [coefficient, 1.2 (0.60-1.83); P < 10(-5)] and BMI [coefficient, 4.1 (0.5-7.8); P = 0.03] but not with all other variables. Optimal cutoff values for S ≥ 1 and S ≥ 2 were 219 dB/m [AUROC, 0.76 (0.67-0.84); sensitivity, 91.1% (78.8-97.5); specificity, 51.6% (38.7-64.2); positive predictive value, 56.9% (44.7-68.6); negative predictive value, 89.2% (74.3-97.0); positive likelihood ratio, 1.88 (1.4-2.5); negative likelihood ratio, 0.17 (0.07-0.5)] and 296 dB/m [AUROC, 0.82 (0.74-0.89); sensitivity, 60.0% (32.3-83.7); specificity, 91.5% (83.9-96.3); positive predictive value, 52.9% (27.8-77.0); negative predictive value, 93.5% (86.3-97.6); positive likelihood ratio, 7.05 (3.2-15.4); negative likelihood ratio, 0.44 (0.2-0.8)], respectively.
Conclusion:
Controlled attenuation parameter could be a useful tool in the clinical management of patients with chronic viral hepatitis for detecting liver steatosis.
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Effect of Hepatic Disease on Pharmacokinetics: Dose Adjustments Due to Hepatic Impairment
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Effect of Hepatic Disease on Pharmacokinetics: Active Drug, Metabolite and Fraction of Metabolized Drug
Hepatic Drug Excretion: Influencing Factors