Accuracy of continuous noninvasive hemoglobin monitoring: a systematic review and meta-analysis

Sang-Hyun Kim1, Marc Lilot, Linda Suk-Ling Murphy

  • 1From the Departments of Anesthesiology & Perioperative Care, Science Library Reference, and Statistics, University of California Irvine, Orange, California.

Insights

Noninvasive hemoglobin (Hb) monitoring devices show a small mean difference compared to lab tests, but wide agreement limits necessitate caution in clinical decisions. Further evaluation of these Hb monitoring tools is recommended.

Area of Science:

  • Clinical Chemistry
  • Medical Devices
  • Biomedical Engineering

Background:

  • Noninvasive hemoglobin (Hb) monitoring devices are increasingly used in clinical settings.
  • The accuracy and precision of these devices against traditional laboratory Hb measurements require rigorous evaluation.
  • Existing literature lacks a systematic review and meta-analysis to consolidate accuracy data.

Purpose of the Study:

  • To systematically review and meta-analyze the accuracy and precision of noninvasive Hb monitoring devices.
  • To compare noninvasive Hb measurements with central laboratory Hb reference standards.
  • To assess the reliability of noninvasive Hb monitoring for clinical decision-making.

Main Methods:

  • A comprehensive literature search was conducted from 2005 to August 2013 across major databases (PubMed, Web of Science, Cochrane Library).
  • Study quality was independently assessed by two reviewers using established guidelines and criteria for method comparison studies.
  • Random-effects meta-analysis was employed to calculate pooled mean difference and standard deviation (SD), including 95% limits of agreement.

Main Results:

  • A total of 32 studies involving 4425 subjects were included in the meta-analysis.
  • The overall pooled mean difference was 0.10 ± 1.37 g/dL, with wide limits of agreement (-2.59 to 2.80 g/dL).
  • Subgroup analysis showed variations, with perioperative settings yielding a mean difference of 0.39 ± 1.32 g/dL and intensive care units -0.51 ± 1.59 g/dL.

Conclusions:

  • While the average difference between noninvasive and laboratory Hb measurements is small, the broad limits of agreement are significant.
  • Clinicians must exercise caution when interpreting noninvasive Hb readings for critical patient management.
  • The current performance of noninvasive Hb monitoring devices warrants careful consideration in clinical practice.
Abstract

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.7K
Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.6K
Holter Monitor: 24-Hour Monitoring01:23

Holter Monitor: 24-Hour Monitoring

Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
5.7K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Pulse Oximetry01:24

Pulse Oximetry

Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
1.6K
Special considerations while measuring oxygen saturation01:19

Special considerations while measuring oxygen saturation

Assessing respiratory rate concurrently with pulse measurement is fundamental to patient care, providing valuable insights into the patient's respiratory function. The normal breathing rate for an adult usually falls within a normal range of 12 to 20 breaths per minute. Abnormal respiratory rates can signal underlying health conditions or the need for immediate intervention.
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
1.0K