Image quality and radiation dose of lower extremity CT angiography using 70 kVp, high pitch acquisition and
Li Qi1, Felix G Meinel2, Chang Sheng Zhou1
1Department of Medical Imaging, Jinling Hospital, Clinical School of South Medical University, Nanjing, Jiangsu, China.
Objectives:
The purpose of this study was to assess image quality and radiation dose of lower extremity CT angiography (CTA) with 70 kVp, high pitch acquisition and sinogram-affirmed iterative reconstruction (SAFIRE).
Methods:
Lower extremity CTAs were performed on 44 patients: 22 patients were examined using protocol A (120 kVp, pitch of 0.85 and 120 ml of contrast agent on a first-generation dual-source CT) (120 kVp group) and 22 patients were evaluated with protocol B (70 kVp, pitch of 2.2 and 80 ml of contrast agent on a second-generation dual-source CT) (70 kVp group). Images from the 120 kVp group were reconstructed with filtered back projection (FBP) and images from the 70 kVp group with SAFIRE. The attenuation, image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Two radiologists subjectively assessed image quality of lower extremity arteries, plantar arterial enhancement and venous contamination of all patients. Radiation dose was compared between the two groups.
Results:
Higher mean intravascular attenuation was obtained in the 70 kVp group (70 vs. 120 kVp group, 555.4 ± 83.4 HU vs. 300.9 ± 81.4 HU, P<0.001), as well as image noise (20.0 ± 2.8 HU vs. 17.5 ± 3.2 HU, P = 0.010), SNR (32.0 ± 7.0 vs. 19.1 ± 6.9, P<0.001) and CNR (28.1 ± 6.6 vs 15.9 ± 6.3, P<0.001). No difference in subjective image quality and plantar arterial enhancement was found between 120 kVp group and 70 kVp group (all P>0.05). The venous contamination score was 1.5 ± 0.8 for 120 kVp group while no venous contamination was found in 70 kVp group. The inter-observer agreement was moderate to good for both groups (0.515∼1, P<0.001). The effective dose was lower in 70 kVp group (0.3 ± 0.1 mSv) than in 120 kVp group (1.6 ± 0.7 mSv)(P<0.001).
Conclusions:
Lower extremity CTA using 70 kVp, high pitch acquisition and SAFIRE, except increasing imaging noise, allows for lower radiation dose and contrast material volume without compromising image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...