Objectives:
Until recently, diagnostic laboratories that wanted to report on the international scale had limited options: they had to align their BCR-ABL1 quantification methods through a sample exchange with a reference laboratory to derive a conversion factor. However, commercial methods calibrated on the World Health Organization genetic reference panel are now available. We report results from a study designed to assess the comparability of the two alignment strategies.
Design And Methods:
Sixty follow-up samples from chronic myeloid leukemia patients were included. Two commercial methods calibrated on the genetic reference panel were compared to two conversion factor methods routinely used at Saint-Louis Hospital, Paris, and at Lille University Hospital. Results were matched against concordance criteria (i.e., obtaining at least two of the three following landmarks: 50, 75 and 90% of the patient samples within a 2-fold, 3-fold and 5-fold range, respectively).
Results:
Out of the 60 samples, more than 32 were available for comparison. Compared to the conversion factor method, the two commercial methods were within a 2-fold, 3-fold and 5-fold range for 53 and 59%, 89 and 88%, 100 and 97%, respectively of the samples analyzed at Saint-Louis. At Lille, results were 45 and 85%, 76 and 97%, 100 and 100%, respectively. Agreements between methods were observed in the four comparisons performed.
Conclusion:
Our data show that the two commercial methods selected are concordant with the conversion factor methods. This study brings the proof of principle that alignment on the international scale using the genetic reference panel is compatible with the patient sample exchange procedure. We believe that these results are particularly important for diagnostic laboratories wishing to adopt commercial methods.
Related Concept Videos
Genetic Material
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Reliability and Validity
mRNA Stability and Gene Expression
Cis-acting Elements involved in mRNA stability
Nonsense-mediated mRNA Decay
Usually, Upf3 binds to an Exon Junction Complex (EJC) at mRNA splice sites. If a ribosome fully translates the mRNA,...
Regulated mRNA Transport