Fast and Accurate Exhaled Breath Ammonia Measurement
Published on: June 11, 2014
The evidence based practice for optimal sample quality for ammonia measurement.
Nora Nikolac1, Jelena Omazic1, Ana-Maria Simundic1
1University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia.
Pre-analytical factors significantly impact ammonia measurement accuracy. Proper sample handling, including immediate cooling and appropriate storage temperature, is crucial for reliable ammonia test results.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Biochemistry
Background:
- Pre-analytical variables are a primary source of variability in laboratory testing.
- Inaccurate laboratory results can lead to incorrect clinical decisions.
- Ammonia measurement is particularly susceptible to pre-analytical influences.
Purpose of the Study:
- To investigate pre-analytical factors affecting ammonia measurement.
- To quantify the impact of these factors on ammonia concentration.
- To identify key variables influencing ammonia assay accuracy.
Main Methods:
- Blood samples from 20 healthy volunteers were analyzed under various pre-analytical conditions.
- Conditions included immediate sample cooling, centrifugation temperature, measurement delay, storage temperature, and tube closure.
- Ammonia levels were measured, alongside alanine aminotransferase (ALT), gamma glutamyltransferase (GGT), and other parameters.
Main Results:
- Cooling samples immediately, centrifugation at 0°C, and storage at 4°C significantly reduced changes in ammonia concentration.
- Observed biases in ammonia levels surpassed clinical acceptance criteria.
- Multiple linear regression identified ALT, GGT, and storage temperature as significant predictors of increased ammonia.
Conclusions:
- Pre-analytical factors introduce significant errors into ammonia measurements.
- Ammonia concentration is most strongly influenced by ALT and GGT activity and storage temperature.
- Optimizing pre-analytical conditions is essential for accurate ammonia testing.
More Related Videos
10:29Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
08:05Measurement of the Potential Rates of Dissimilatory Nitrate Reduction to Ammonium Based on 14NH4+/15NH4+ Analyses via Sequential Conversion to N2O
Published on: October 7, 2020
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Titration of a Weak Base with a Strong Acid
Using the ICE table and substituting the Kb value, we calculate the initial pH of 50 mL of 0.1 M ammonia to be 11.11. Addition of 25 mL of 0.1 M hydrochloric acid to this solution of ammonia results in a buffer with an equal concentration of ammonia and ammonium ions. The pH of this buffer can be calculated by substituting these...