Practical limitations of two devices used for the measurement of sub-bandage pressure: implications for clinical
1BPharm, PhD, Medical Device Consultancy, MedetecTM.
Objective:
This study was undertaken to examine the accuracy of two hand-held devices commonly used to measure the pressures produced by extensible bandages.
Method:
The performance of the pneumatic sensors of two devices, the Kikuhime and Picopress instruments was first examined in air in a compression chamber, then subsequently beneath multiple layers of bandages applied to standard cylinders with predetermined levels of tension.
Results:
In the compression chamber, both instruments provided readings that were typically within 1 mmHg of the reference value, but on curved formers in free air or beneath bandages the accuracy of both sensors was greatly reduced, influenced both by the curvature of the cylinders and the volume of air contained in the sensor capsule. The Picopress instrument recorded pressures up to 70% higher than predicted, particularly on the smaller cylinders and at the lower end of the pressure range (circa 20mmHg). At 40mmHg measured pressure were around 40% higher than predicted values in some instances. The accuracy of the Kikuhime was greatly influenced by the calibration technique but percentage deviations as high as 150% were recorded in some tests.
Conclusion:
Pneumatic pressure sensors used to record the pressures developed beneath compression bandages are much less accurate than is commonly believed. Calibration studies using air chambers or water tanks have no relevance to the normal clinical use of these devices. These limitations should be considered by clinicians when making judgments about the performance of other medical devices such as bandages or stockings. They also call into question the validity of many published studies which rely upon such measurements.
Declaration Of Interest:
This project was generously supported by 3M Healthcare who provided a small research grant to assist with the costs associated with the construction of the test equipment.
More Related Videos
06:45Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
Published on: May 2, 2025
04:34A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: