Noise from total hip replacements: a case-controlled study
M C Wyatt1, S Jesani1, C Frampton2
1Wellington Regional Hospital, Riddiford Street, Newtown, Wellington 6021, New Zealand.
Objectives:
Our study aimed to examine not only the incidence but also the impact of noise from two types of total hip replacement articulations: ceramic-on-ceramic and ceramic-on-polyethylene.
Methods:
We performed a case-controlled study comparing subjective and objective questionnaire scores of patients receiving a ceramic-on-ceramic or a ceramic-on-polyethylene total hip replacement by a single surgeon.
Results:
There was a threefold higher incidence of noise from patients in the ceramic-on-ceramic group compared with the control group. The impact of this noise was significant for patients both subjectively and objectively.
Conclusions:
This study reports a high patient impact of noise from ceramic-on-ceramic total hip replacements. This has led to a change in practice by the principal author. Cite this article: Bone Joint Res 2014;3:183-6.