Related Experiment Video
Updated: Apr 28, 2026

Dual-Dye Optical Mapping of Hearts from RyR2R2474S Knock-In Mice of Catecholaminergic Polymorphic Ventricular Tachycardia
Published on: December 22, 2023
Inappropriate Detection of a Supraventricular Tachycardia as Dual Tachycardia by the PR Logic™ Algorithm
Ajit Thachil1, Sridevi Chennapragada2, Narasimhan Calambur2
1Lisie Hospital, Kochi, Kerala, India.
Abstract:
Tachycardia detection and therapy algorithms in Implantable Cardioverter-Defibrillators (ICD) reduce, but do not eliminate inappropriate ICD shocks. Awareness of the pros and cons of a particular algorithm helps to predict its utility in specific situations. We report a case where PR logic™, an algorithm commonly used in currently implanted ICDs to differentiate supraventricular tachycardia (SVT) from ventricular tachycardia resulted in inappropriate detection and shock for an SVT, and discuss several solutions to the problem.