The design and operation of a novel dual-laser excitation Raman instrument is described. The use of two lasers of differing wavelengths allows for a Raman spectrum covering all fundamental modes of vibration to be collected while minimizing fluorescence and allowing for spatial compression of the spectrum on an imaging detector. The use of diode lasers with integrated distributed Bragg reflector gratings facilitates the use of an integrated thermoelectric cooler to allow collection of shifted excitation spectra for both of the lasers, further enhancing the rejection of fluorescence. An example is given, which uses seven excitation wavelengths for each laser to reconstruct the Raman spectrum of a solvent in the presence of a highly fluorescent dye by using a sequentially shifted excitation Raman reconstruction algorithm.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...